Advisor(s)
Abstract(s)
An ofloxacin-selective electrode based on incorporation of a ofloxacin-tetrakis[3,5-
bis(trifluoromethyl)phenyl]borate ion-pair complex in a poly(vinyl chloride) membrane was
developed. The sensor displayed a Nernstian response over a wide concentration range
(7×10-6–1×10-2 mol L-1) with a slope of 54.9 ± 1.3 mV per decade of concentration, a detection
limit of 4×10-6 mol L-1 and adequate selectivity toward some inorganic and organic species.
The sensor can be used for a period of more than 8 months without significant change
in its characteristics. Desenvolveu-se um eléctrodo selectivo à ofloxacina, baseado no par iónico ofloxacina-
-tetrakis[3,5-bis(trifluorometil)fenil]borato incorporado numa membrana de cloreto de
polivinilo. O sensor apresenta uma resposta Nernstiana num alargado intervalo de concentrações (7×10-6–1×10-2 mol L-1), com um declive de 54,9±1,3 mV por década de concentração,
um limite de detecção de 4×10-6 mol L-1 e adequada selectividade relativamente a
algumas espécies orgânicas e inorgânicas. O seu tempo de vida é superior a oito meses, não
havendo alteração das suas características durante este período.
Description
Keywords
Fluoroquinolones Ofloxacin Ion-selective electrodes Potentiometry Ion-pair sensor PVC membrane Fluoroquinolonas Ofloxacina Eléctrodos selectivos de ião Potenciometria Sensor par-iónico Membrana de PVC
Citation
Revista da Faculdade de Ciências da Saúde. Porto: Edições Universidade Fernando Pessoa. ISSN 1646-0480. 7 (2010) 28-41.