Souto, RenataPimenta, AdrianaLeal, FernandaCatarino, RitaSantos, GoryTavares, Tânia2012-04-262012-04-262010Revista da Faculdade de Ciências da Saúde. Porto: Edições Universidade Fernando Pessoa. ISSN 1646-0480. 7 (2010) 28-41.1646-0480http://hdl.handle.net/10284/3010An ofloxacin-selective electrode based on incorporation of a ofloxacin-tetrakis[3,5- bis(trifluoromethyl)phenyl]borate ion-pair complex in a poly(vinyl chloride) membrane was developed. The sensor displayed a Nernstian response over a wide concentration range (7×10-6–1×10-2 mol L-1) with a slope of 54.9 ± 1.3 mV per decade of concentration, a detection limit of 4×10-6 mol L-1 and adequate selectivity toward some inorganic and organic species. The sensor can be used for a period of more than 8 months without significant change in its characteristics. Desenvolveu-se um eléctrodo selectivo à ofloxacina, baseado no par iónico ofloxacina- -tetrakis[3,5-bis(trifluorometil)fenil]borato incorporado numa membrana de cloreto de polivinilo. O sensor apresenta uma resposta Nernstiana num alargado intervalo de concentrações (7×10-6–1×10-2 mol L-1), com um declive de 54,9±1,3 mV por década de concentração, um limite de detecção de 4×10-6 mol L-1 e adequada selectividade relativamente a algumas espécies orgânicas e inorgânicas. O seu tempo de vida é superior a oito meses, não havendo alteração das suas características durante este período.engFluoroquinolonesOfloxacinIon-selective electrodesPotentiometryIon-pair sensorPVC membraneFluoroquinolonasOfloxacinaEléctrodos selectivos de iãoPotenciometriaSensor par-iónicoMembrana de PVCDevelopment and analytical characterization of ofloxacin ion-selective electrodesjournal article